4 Remote Interface Reference
282 Keysight 33210A User’s Guide
Calibration Commands
For an overview of the calibration features of the function generator, refer to
“Calibration Overview” in Chapter 3 starting on page 155. For a detailed
discussion of the function generator’s calibration procedures, refer to Chapter 4 in
the Keysight 33210A Service Guide.
CAL:SECure:STATe {OFF|ON},<code>
CAL:SECure:STATe?
Unsecure or secure the instrument for calibration. The calibration code may
contain up to 12 characters. The :STAT? query returns “0” (OFF) or “1” (ON).
CAL:SETup <0|1|2|3| . . . |75>
CAL:SETup?
Configure the function generator’s internal state for each of the calibration steps
to be performed. The :SET? query reads the calibration setup number and returns
a value from “0” to “75”.
CAL:VALue <value>
CAL:VALue?
Specify the value of the known calibration signal as outlined in the calibration
procedures in the Keysight 33210A Service Guide. Use the CAL:SET command to
configure the function generator’s internal state for each of the calibration steps
to be performed. The :VAL? query returns a number in the form
“+1.0000000000000E+01”.
CAL?
Perform a calibration of the instrument using the specified calibration value
(CAL:VAL command). Before you can calibrate the function generator, you must
unsecure it by entering the correct security code. Returns “0” (PASS) or “1” (FAIL).