EasyManua.ls Logo

Keysight Technologies 33210A - Carrier Frequency

Keysight Technologies 33210A
367 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Features and Functions
94 Keysight 33210A User’s Guide
Carrier Frequency
The maximum carrier frequency depends on the function selected as shown
below. The default is 1 kHz for all functions.
–The carrier frequency must always be greater than or equal to the frequency
deviation. If you attempt to set the deviation to a value greater than the carrier
frequency (with FM enabled), the function generator will automatically adjust
the deviation to the maximum value allowed with the present carrier
frequency.
–The sum of the carrier frequency and deviation must be less than or
equal to
the maximum frequency for the selected function
plus 100 kHz
(10.1 MHz for
sine and square waveforms). If you attempt to set the deviation to a value that
is not valid, the function generator will automatically adjust it to the maximum
value allowed with the present carrier frequency.
Front-Panel Operation: To set the carrier frequency, press the Freq softkey for
the selected function. Then use the knob or numeric keypad to enter the
desired frequency.
Remote Interface Operation:
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Function Minimum Frequency Maximum Frequency
Sine
Square
1 mHz
1 mHz
10 MHz
10 MHz

Table of Contents

Other manuals for Keysight Technologies 33210A