3 Features and Functions
118 Keysight 33210A User’s Guide
Burst Phase
The burst phase defines the starting phase of the burst.
Burst phase: -360 degrees to +360 degrees. The default is 0 degrees.
From the remote interface, you can set the starting phase in degrees or radians
using the UNIT:ANGL command (see page 170).
From the front panel, the starting phase is always displayed in degrees
(radians
are not available). If you set the starting phase in radians from
the remote
interface and then return to front-panel operation, you will
see that the function
generator converts the phase to degrees.
For sine, square, and ramp waveforms, 0 degrees is the point at which the
waveform crosses zero volts (or the dc offset value) in a positive-going
direction.
The burst phase is also used in the gated burst mode. Then the gate signal
goes false, the current waveform cycle is completed and then the function
generator stops. The output will remain at the voltage level corresponding to
the starting burst phase.
Front-Panel Operation: To set the burst phase, press the Start Phase softkey
and then use the knob or numeric keypad to enter the desired phase in
degrees.
Remote Interface Operation:
BURSt:PHASe {<angle>|MINimum|MAXimum}