Remote Interface Reference 4
Keysight 33210A User’s Guide 281
Miscellaneous Status Register Commands
*CLS
Clear the event register in all register groups. This command also clears the error
queue and cancels a *OPC operation.
STATus:PRESet
Clear all bits in the Questionable Data enable register and the Standard Operation
enable register.
*PSC {0|1}
*PSC?
Power-On Status Clear. Clear the Standard Event enable register and Status Byte
condition register at power on (*PSC 1). When *PSC 0 is in effect, these two
registers are not cleared at power on. The default is *PSC 1. The *PSC? query
returns the power-on status clear setting. Returns “0” (do not clear at power on)
or “1” (clear at power on).
*OPC
Set the “Operation Complete” bit (bit 0) in the Standard Event register after the
previous commands have completed. When used with a bus-triggered sweep or
burst, you may have the opportunity to execute commands after the *OPC
command and before the “Operation Complete” bit is set in the register.