4 Remote Interface Reference
222 Keysight 33210A User’s Guide
4 Set the sweep time.
Use the SWE:TIME command to set the number of seconds required to sweep
from the start frequency to the stop frequency.
5 Select the sweep trigger source.
Use the TRIG:SOUR command to select the source from which the sweep will
be triggered.
6 Set the marker frequency. (Optional)
If desired, you can set the frequency at which the signal on the front-panel
Sync connector goes to a logic low during the sweep. Use the MARK:FREQ
command to set the marker frequency to any value between the start
frequency and stop frequency. Use the MARK ON command to enable the
frequency marker.
7 Enable the sweep mode.
After you have set up the other sweep parameters, use the SWE:STAT ON
command to enable the sweep mode.