3 Features and Functions
114 Keysight 33210A User’s Guide
–When the gated mode is selected, the burst count, burst period, and trigger
source are ignored (these parameters are used for the triggered burst mode
only). If a manual trigger is received, it is ignored, and no error will be
generated.
–When the gated mode is selected, you can also select the polarity of the signal
on the rear-panel Trig In connector.
Front-Panel Operation: After enabling bursts, press the N Cycle (triggered) or
Gated softkey.
To select the polarity of the external gate signal on the Trig In connector, press
the Polarity softkey. The default polarity is POS (true-high logic).
Remote Interface Operation:
BURSt:MODE {TRIGgered|GATed}
Use the following command to select the polarity of the external gate signal on
the Trig In connector. The default if NORM (true-high logic).
BURST:GATE:POLarity {NORMal|INVerted}