3 Features and Functions
156 Keysight 33210A User’s Guide
To Unsecure for Calibration
You can unsecure the function generator either from the front panel or over the
remote interface. The function generator is secured when shipped from the
factory and the security code is set to “AT33210A”.
– Once you enter a security code, that code must be used for both front-
panel
and remote operation. For example, if you secure the function generator
from
the front panel, you must use that same code to unsecure it from the remote
interface.
– Front-Panel Operation: Press and then select the Secure Off softkey
from the “Test / Cal” menu.
– Remote Interface Operation: To unsecure the function generator, send the
following command with the correct security code.
CAL:SECURE:STATE OFF,AT33210A
To Secure Against Calibration
You can secure the function generato
r either from the front panel or over the
remote interface. The function generator is secured when shipped from the factory
and the security code is set to “AT33210A”.
– Once you enter a security code, that code must be used for both front-
panel
and remote operation. For example, if you secure the function generator
from
the front panel, you must use that same code to unsecure it from the remote
interface.
– Front-Panel Operation: Press and then select the Secure On softkey
from the “Test / Cal” menu.
– Remote Interface Operation: To secure the function generator, send the
following command with the correct security code.
CAL:SECURE:STATE ON,AT33210A