4 Remote Interface Reference
246 Keysight 33210A User’s Guide
If you copy to a waveform name that already exists, the previous waveform is
overwritten (and no error will be generated). However, you cannot overwrite
any of the five built-in waveforms.
Up to four user-defined waveforms can be stored in non-volatile memory. If
memory is full and you try to copy a new waveform to non-volatile memory, a
“Not enough memory” error is generated. Use the DATA:DEL command to
delete the waveform in volatile memory or any of the four user-defined
waveforms in non-volatile memory. Use the DATA:CAT? command to list all
waveforms currently stored in volatile and non-volatile memory. The default
selection is EXP_RISE.
The following statement shows how to use the DATA:COPY command to copy
the VOLATILE waveform into named storage “ARB_1”.
DATA:COPY ARB_1, VOLATILE
FUNCtion:USER {<arb name>|VOLATILE}
FUNCtion:USER?
Select one of the five built-in arbitrary waveforms, one of four user-defined
waveforms, or the waveform currently downloaded to volatile memory. The
:USER? query returns “EXP_RISE”, “EXP_FALL”, “NEG_RAMP”, “SINC”,
“CARDIAC”, “VOLATILE”, or the name of any user-defined waveforms in
non-volatile memory. The default selection is “EXP_RISE”.
Note that this command does not output the selected arbitrary waveform.
Use the FUNC USER command (see the following page) to output the selected
waveform.
The names of the five built-in arbitrary waveforms are:
“EXP_RISE”, “EXP_FALL”,
“NEG_RAMP”,
“SINC”, and “CARDIAC”.
To select the waveform currently stored in volatile memory, specify the
VOLATILE parameter. The keyword “VOLATILE” does not have a short form.
If you select a waveform name that is not currently downloaded, a “Specified
arb waveform does not exist” error is generated.