Loading...
4 Remote Interface Reference
258 Keysight 33210A User’s Guide
*TST?
Perform a complete self-test of the function generator. Returns “+0” (PASS) or
“+1” (FAIL). If the test fails, one or more error messages will be generated to
provide additional information on the failure. Use the SYST:ERR? command to
read the error queue (see page 255).
SYSTem:VERSion?
Query the function generator to determine the present SCPI version. Returns a
string in the form “YYYY.V”, where “YYYY” represents the year of the version, and
“V” represents a version number for that year (e.g., 1999.0).
SYSTem:BEEPer
Issue a single beep immediately.
SYSTem:BEEPer:STATe {OFF|ON}
SYSTem:BEEPer:STATe?
Disable or enable the tone heard when an error is generated from the front-panel
or over the remote interface. The current selection is stored in non-volatile
memory. The :STAT? query returns “0” (OFF) or “1” (ON).
SYSTem:KLOCk[:STATe] {OFF|ON}
Set front panel keyboard lock OFF (the default) or ON. With :KLOC ON, the front
panel keyboard is locked, including the key unless it has been excluded. To
lock the keyboard without locking the key, send SYST:KLOC:EXCL LOC first,
before SYST:KLOC ON.
SYSTem:KLOCk:EXCLude {NONE|LOCal}
SYSTem:KLOCk:EXCLude?
Setting :EXCL NONE (the default) sets no exclusion, and SYST:KLOC ON locks
the entire keyboard including the key.
Setting :EXCL LOC excludes the key, and SYST:KLOC ON locks the
keyboard, except the key.