4 Remote Interface Reference
264 Keysight 33210A User’s Guide
SYSTem:COMMunicate:LAN:NETBios {OFF|0|ON|1}
SYSTem:COMMunicate:LAN:NETBios?
Disables or enables the use of NETBios to automatically assign an IP address to
the 33210A for peer-to-peer naming on a private network that uses the NETBios
naming service. The factory setting is enabled (ON). The query returns “0” (OFF) or
“1” (ON).
If you change the NETBios setting, you must cycle power on the instrument to
activate the new setting.
The NETBios setting is stored in non-volatile memory, and does not change
when power has been off, or after an instrument reset (*RST).
SYSTem:COMMunicate:LAN:TELNet:PROMpt <string>
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when using a Telnet session to communicate with
the 33210A. The default prompt is “33210A>”.
The 33210A uses LAN port 5024 for Telnet sessions.
Telnet session can typically be started as follows from a host computer shell:
telnet <IP_address> <port>
–For example:
telnet 169.254.2.20 5024
To exit a Telnet session, press <Cntr-D>.
The prompt is a quoted string of up to 15 characters.
The command prompt is stored in non-volatile memory, and does not change
when power has been off, or after an instrument reset (*RST).