Remote Interface Reference 4
Keysight 33210A User’s Guide 271
The Status Byte condition register is cleared when:
You execute the *CLS (clear status) command.
You read the event register from one of the other register groups (only the
corresponding bits are cleared in the condition register).
The Status Byte enable register is cleared when:
You execute the *SRE 0 command.
You turn the power on and have previously configured the function generator
to clear the enable register using the *PSC 1 command. Note that the enable
register will not be cleared at power-on if you have configured the function
generator using the *PSC 0 command.