5 Error Messages
296 Keysight 33210A User’s Guide
-131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have
misspelled the suffix. Example: SWE:TIME 0.5 SECS
-138
Suffix not allowed
A suffix is not supported for this command.
Example
:
BURS:NCYC 12 CYC
-148
Character data not allowed
A discrete parameter was received but a character string or numeric parameter
was expected. Check the list of parameters to verify that you have used a valid
parameter type. Example: DISP:TEXT ON
-151
Invalid string data
An invalid character string was received. Check to see if you have enclosed the
character string in quotation marks and verify that the string contains valid ASCII
characters. Example: DISP:TEXT 'TESTING (the ending quote is missing)
-158
String data not allowed
A character string was received but is not allowed for this command. Check the
list of parameters to verify that you have used a valid parameter type.
Example: BURS:NCYC 'TEN'
-161
Invalid block data
Applies only to the
DATA:DAC VOLATILE
command
.
For a definite-length block, the number of bytes of data sent does not match the
number of bytes that you specified in the block header. For an indefinite-length
block, an EOI (End-or-Identify) was received without an accompanying <new line>