EasyManua.ls Logo

Keysight 81150A - Voltage Limits

Keysight 81150A
672 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features and Functions
81150A and 81160A User’s Guide 115
3.2.11 Voltage Limits
Introduction
The voltage limits, limits the range that can be set by software but does not
apply some kind of programmable hardware limits in the amplifier.
Charactersitics
Use the voltage limits for the level parameters to prevent accidental damage
of the device under test.
After you switch on the limits, the level parameters on the function screens
cannot be adjusted outside the range specified as voltage limits if the output
is switched on.
Front Panel
Operation
Press the key, Output Setup and then Voltage Limits softkey. To
enable/disable the Voltage Limits, press the Voltage Limits softkey again.
High Volt Limit is used to set the maximum allowed voltage.
Low Volt Limit is used to set the minimum allowed voltage.
Remote Interface
Operation
:VOLTage[1|2]:LIMit:STATe {OFF|ON}
:VOLTage[1|2]:LIMit{<voltage>}MINimum|MAXimum}
:VOLTage[1|2]LIMit:LOW {<voltage>|MINimum|MAXimum}

Table of Contents