3.9.6 Phase Deviation
Introduction
The phase deviation setting represents the peak variation in phase of the
modulated waveform from the carrier waveform. The phase deviation can be
set from 0 to 360 degrees. The default is 180 degrees.
Since a 360 degree phase deviation is equivalent to 0 degrees, the maximum
effective deviation setting is 180 degrees.
Front Panel
Operation
After enabling PM, press the Phase Deviation softkey. Then use the knob or
numeric keypad to enter the desired deviation.
Remote Interface
Operation
:PM[1|2]:DEViation{<deviation
in degrees>|MINimum|MAXimum}