Features and Functions
81150A and 81160A User’s Guide 127
Remote Interface
Operation
:OUTPut[1|2]:TRIGger:VOLTage {TTL|ECL}
3.2.19 Sync Output
Introduction
The Sync Outputs exist on the 81160A, only. The 81150A has one Trigger
Output per channel and one Strobe Outputs per channel.
For the one channel instrument as well as the two channel instrument of
81160A two high speed Sync Out A and Sync Out B outputs are provided at
the front panel.
Mode of Operation
The Sync Output signals can be configured very flexible by an internal
switch matrix to output the ‘Logical Trigger Signal’ functionality or ‘Logical
Strobe Signal’ functionality according to the following switch matrix:
81160A-001 (1 channel instrument)
Output at BNC Connector
Internal Logical Signal
Sync Out A source:
None, Logical Trigger Signal Channel 1,
Logical Strobe Signal Channel 1
Sync Out B source:
None, Logical Trigger Signal Channel 1,
Logical Strobe Signal Channel 1
Note: It is e.g. possible, that the Logical Trigger Signal 1 functionality is
routed simultaneously to Sync Out A and Sync Out B.
81160A-002 (2 channel instrument)
Output at BNC Connector
Internal Logical Signal
Sync Out A source:
None, Logical Trigger Signal Channel 1,
Logical Strobe Signal Channel 1, Logical
Trigger Signal Channel 2, Logical Strobe
Signal Channel 2
Sync Out B source:
None, Logical Trigger Signal Channel 1,
Logical Strobe Signal Channel 1, Logical
Trigger Signal Channel 2, Logical Strobe
Signal Channel 2
Note: It is e.g. possible, that the Logical Trigger Signal Channel 1
functionality is routed simultaneously to Sync Out A and Sync Out B.