The duty cycle setting does not apply to a square waveform used as the
modulating waveform for AM, FM, PM or PWM. A 50% duty cycle is always
used for a modulating square waveform. The duty cycle setting applies only
to a square waveform carrier.
After selecting the square wave function, press the Duty Cycle softkey.
Then use the knob or numeric keypad to enter the desired duty cycle.
Remote Interface
Operation
The following function is used to configure the duty cycle remotely:
:FUNCtion[1|2]:SQUare:DCYCl{<percentage>|
MINimum|MAXimum}