EasyManua.ls Logo

Keysight 81150A - Page 327

Keysight 81150A
672 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Programming Reference
81150A and 81160A User’s Guide 327
Command
:APPL[1|2]:SQU
Long
:APPLy[1|2]:SQUare
Parameters
[<frequency>[,<amplitude>[,<offset>]]]
Parameter Suffix
Description
Output a square wave with the specified frequency, amplitude, and dc offset.
This command overrides the current duty cycle setting and automatically
selects 50%. The waveform is output as soon as the command is executed.
Example
:APPL1:SQU 110e6, 2.4, -1.25