EasyManua.ls Logo

Keysight 81150A - Page 446

Keysight 81150A
672 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
446
Command
:FUNC[1|2]:PULS:WIDT[?]
Long
[:SOURce]:FUNCtion[1|2]:PULSe:WIDTh[?]
Parameters
{<NR3>|MINimum|MAXimum}
Parameter Suffix
S
Description
Set the pulse width in seconds. The pulse width represents the time from
the start of rising edge of the pulse to the start of the next falling edge. The
:FUNC[1|2]:PULS:WIDT? query returns the pulse width in seconds.
This command is affected by the :FUNC[1|2]:PULS:HOLD
command, which determines the value to be held constant as the period is
adjusted: the specified pulse width value, or the specified pulse duty cycle
value, or the specifies trailing delay. See the :FUNC[1|2]:PULS:HOLD
command for further information.
Example
:FUNC2:PULS:WIDT 21.5NS