Remote Programming Reference
81150A and 81160A User’s Guide 519
Command
:CAL?
Long
:CALibration[:ALL]?
Parameters
Parameter Suffix
Description
Disconnect all cables before executing calibration. Self-calibration includes
the following:
Levels at normal and inverted outputs
Levels at Trigger-Out
Levels at Strobe-Out
Sensitivity of External-In
Input voltage window and OV level at Modulation-In
Example
:CAL?
Response: 1