Features and Functions
81150A and 81160A User’s Guide 119
3.2.14 Output Control
You can disable or enable the front-panel Output connector. By default, the
output is disabled at power on to protect other equipment. When enabled,
the Output key is illuminated.
Output Control
characteristics
If an excessive external voltage is applied to the front-panel Output
connector, an error message will be displayed and the output will be
disabled. To re-enable the output, remove the overload from the Output
connector and press both the Output keys (for normal and inverted output)
for the respective channel to enable the output.
Press both, the normal and the inverted Output keys to enable or disable the
output.
Remote Interface
Operation
The following function is used to configure the output control remotely:
:OUTput[1|2] {0|1|OFF|ON}
:OUTPut[1|2]:COMPlement {0|1|OFF|ON}
The APPLy command overrides the current setting and automatically
enables the Output connector.
In case of an overload condition, both outputs of the channel (Normal and
Inverted) will be turned off.