Features and Functions
81150A and 81160A User’s Guide 145
3.4.3 Leading Edge/Trailing Edge
The edge time sets each edge transition time (rising and falling) of the
pulse. For each transition, the edge time represents the time from the 10%
threshold to the 90% threshold.
81150A: Edge time: 2.5 ns to 1000 s (see restrictions below). The
default edge time is 2.5 ns.
81150A: The minimum edge time is limited to 7.5 ns. If the amplifier
type is set to “max. Amplitude.”
81160A: Edge time: 1.0 ns to 1000 s. The default edge time is 1.0 ns.
Lead Edge and Trail Edge may be specified in seconds or % of period.
Trail Edge can be configured to follow the leading edge.
After selecting the pulse function, press the Lead Edge/Trail Edge softkeys.
Then use the knob or numeric keypad to enter the desired edge time.