Remote Programming Reference
81150A and 81160A User’s Guide 335
:DATA[1|2]:ATTRibute:PTPeak?
Query the peak-to-peak value of all data points for the specified arbitrary
waveform. The default arb- name is the arbitrary waveform currently active
(selected with FUNC:USER command). This command returns a value from
“0” to “+1.0”, with “+1.0” indicating full amplitude available.
The maximum amplitude will be limited if the data points do not span the
full range of the output DAC (Digital-to-Analog Converter). For example, the
built-in “Sinc” waveform does not use the full range of binary values
between ±1 and therefore its maximum amplitude is lower (into 50 ohms).
If you query a waveform that is not currently stored in memory, a “Specified
arb waveform does not exist” error is generated.
:DATA:ATTR:PTP? EXP_RISE
Response
1.0000E+00