Appendix
76
Start Phase
Defines the start phase of the waveform that is bursted. The start phase is
only available for sinewave, and arb waveforms. Noise and DC do not have a
period and thus no phase information. Square, ramp and pulse is generated
in a way that does not allow the use of a start phase.
The Start Phase can vary between -360 to 360.
Set the
period/frequency
Steps:
Set the period/frequency of the sinewave to 2 ms.
Enable triggered mode.
Select Internal trigger source.
Set the internal frequency to 50 Hz or the internal period to 20 ms.