EasyManua.ls Logo

Keysight E4981A - Page 234

Keysight E4981A
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
234 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Error Messages
Error Messages (Alphabetical Order)
-
C
41 Correction Measurement Aborted
This error occurs when the correction data measurement is aborted.
1200 CPU bd FLASH ROM write error
An error is generated while writing data to FLASH.
Contact Keysight Technology’s Sales and Service Office or the company from
which you purchased the device.
1201 CPU bd EEPROM write error
An error is generated while writing data to EEPROM.
Contact Keysight Technology’s Sales and Service Office or the company from
which you purchased the device.
-100 Command error
A comprehensive syntax error has occurred for which the E4981A cannot
detect further details of the error. This error code simply indicates the
occurrence of a command error that is defined in IEEE488.2,11.5.1.1.4.
-140 Character data error
An error not included in the error numbers between -141 and -149 has
occurred during the syntax analysis of a character data element.
-148 Character data not allowed
Character data not allowed for this operation.
D
-104 Data type error
The parser has recognized impossible data elements. For example, numeric
value or string data is expected, but block data is sent.
-222 Data out of range
A data element (that does not violate the standard) has been received out of
the range defined for the E4981A.
-230 Data corrupt or stale
The data is invalid or a newly initiated read operation has not been completed
since the latest access.

Table of Contents

Other manuals for Keysight E4981A

Related product manuals