Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 237
Error Messages
Error Messages (Alphabetical Order)
-224 Illegal parameter value
The parameter recieved is not correct. For example, though a correct program
message was “:CALC1:FORM CP,” a wrong program message, “:CALC1:FORM
RP,” was received.
L
46 LOAD measurement incomplete
This error occurs when the cable correction LOAD measurement is incomplete.
M
43 Measurement failed
A measurement failure has occurred during measuring the correction data.
-109 Missing parameter
The number of parameters is less than required by the command. For example,
although the :CREJ:LIM command requires one parameter such as
:CREJ:LIM 3”, no parameter is added to give “:CREJ:LIM”.
-250 Mass storage error
An error occurred while accessing the external mass storage device.
N
83 No data to load
There is no setup data for the selected number or no external mass storage
device is connected.
-120 Numeric data error
Numeric data is improper.
-128 Numeric data not allowed
A numeric value data element (that does not violate the standard) has been
received where the E4981A does not accept any numeric value data element.