Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 41
Getting Started
Learning Basic Measurement Procedure
Figure 3-4 Secondary parameter selection menu screen
Setting up measurement signal frequency
Set up the frequency of the signal applied to the DUT (capacitor) during
measurement.
Step 1. Press the [Display Format] key.
Step 2. Press the MEAS DISPLAY softkey.
Step 3. Use the cursor key to select the FREQ field.
Step 4. Use the softkeys or entry keys to enter the frequency. When data is entered
with the numeric entry keys, the softkeys change to unit labels (Hz, kHz, MHz)
for Option 001 and (Hz, kHz) for Option 002.
Figure 3-5 Measurement signal frequency selection menu screen
In Option 002, 1 MHz frequency is not available.