48 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Getting Started
Learning Basic Measurement Procedure
3-
Description of softkeys
To enable/disable or otherwise control the behavior of short correction, use the
following softkeys:
If SHORT impedance |Zs| (= ) is greater than 20 Ω (an unsuitable level
for the data used in SHORT correction), the warning message “Out of limit”
appears on the System message area on the screen
If a measurement failure occurs while measuring the data for correction, an error
message appears on the screen.
Softkey Description
ON Enables short correction.
OFF Disables short correction.
MEAS
SHORT
Starts short correction.
Even if this warning message appears, the data for SHORT correction is still
used. However, you should recheck the connection between the test fixture
and the UNKNOWN terminal and confirm that the SHORT correction
procedure was done correctly.
If an error occurs, the data for correction before the measurement remains
without change.