Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 49
Getting Started
Learning Basic Measurement Procedure
Checking data for SHORT correction
You can check the measured SHORT impedance value (the data for SHORT
correction). The procedure is given below.
Step 1. Press the [Meas Setup] key.
Step 2. Press the CORRECTION softkey.
Step 3. Use the cursor keys to select the SHORT R-X field (as shown in Figure 3-12).
Figure 3-12 SHORT correction data
Step 4. Use the following softkeys:
Softkey Description
R-X Displays the measured R-X value on the screen.
Ls-Rs Displays the measured Ls-Rs value on the screen.
For information on measuring data for Load correction, refer to “Obtaining
(measuring) data for LOAD correction” on page 102.
For checking data for Load correction, refer to “Checking
displaying/setting up data for LOAD correction” on page 111.