Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 87
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
Data B
Outputs the measurement data for the secondary parameter.
The Data B field uses the following two different fixed-length ASCII formats:
Status
Represents the measurement result status by using one of the following status
values:
The Status field uses the following two-character fixed-length ASCII format:
SN (S:+/-, N:0 to 2)
Bin no. (IN/OUT)
Represents the bin sorting results as well as IN/OUT evaluation results, as
shown below:
When the “effective digits”
setting is off:
(:FORMat:ASCii:LONG OFF)
SN.NNNNNESNN
(S:+/-, N:0 to 5, E: exponent character)
When the “effective digits”
setting is on:
(:FORMat:ASCii:LONG ON)
SN.NNNNNNNNNESNN
(S:+/-, N:0 to 9, E: exponent character)
0 Measurement successfully completed.
+1 Overload.
+2 Low C or No contact.
If the value is 1, the measurement data is 9.9E37; if the value is 0 or 2, the
actual measurement data is output.
The data buffer memory contains a measurement result with no data (as
represented by the status value of –1), but it is excluded when the data is
saved into USB memory.
0OUT_OF_BINS
+1 to +9 BIN 1 through BIN 9
+10 AUX_BIN
+11 BIN_NA