88 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Setting up Measurement Conditions and Display
Saving/Recalling Instrument Setup State (Save/Recall Function)
4-
The <BIN No.> data output format is either 2- or 3-character fixed-length
ASCII format:
SN or SNN (S:+/-, N:0 to 9)
Example 4-1 Example of saved measurement result data
+1.059517689E-24,+1.954963777E+00,+0,+0
+9.706803904E-25,+2.095857894E-01,+0,+0
+2.172725184E-24,+2.072965495E-01,+0,+0
+3.660460872E-25,+7.172688291E+00,+0,+0
+1.135428381E-24,+6.490636201E-01,+0,+0
+1.384790632E-24,+2.193020669E+00,+0,+0
+3.829879310E-26,+2.788435221E+01,+0,+0
To save measurement results into USB memory:
Step 1. Plug a USB memory stick into the front USB port.
Step 2. Press [Save/Recall].
Step 3. Press the SAVE DATA softkey.
Step 4. Press the START LOG softkey and then press the following softkeys to enter the
measurement results into the data buffer memory.
Step 5. Start measurement. The data buffer memory is filled with up to 1000 sets of
measurement results.
Step 6. Press the SAVE & STOP softkey to save results into the USB memory.
Step 7. When the data has been saved into the USB memory, a “Storing data
completed. : E498xXXX.csv” message appears in the system message area.
Softkey Description
START LOG Starts logging the measurement results into the data buffer memory.
SAVE & STOP Copies the data from the data buffer memory into the USB memory.
Then stops saving the measurement results into the data buffer memory
and clears the data buffer memory.
The above mentioned steps stores the data in buffer 3, enabling to save the
data to USB memory.