PWR-01 143
Appendix Troubleshooting
Unable to control functions properly
The fan speed is fast
Symptom Items to check Remedy
The bleeder circuit on/off
function does not work.
Are you using master-slave parallel
operation or series operation?
When using master-slave parallel operation or
series operation, set the bleeder circuit settings
(CF01) of all connected units to the same
value.
p.58
Symptom Items to check Remedy
The fan rotates faster than
normal when the POWER
switch is turned on.
Is the bleeder circuit set to hyper bleeder
(CF01: HYP)?
If hyper bleeder (CF01: HYP) was specified the
last time the POWER switch was turned off, the
fan speed will be set to the fastest setting. If
necessary, set the bleeder circuit (CF01).
p.58
p.76
See
See