PWR-01 129
Specifications | Common specifications
Signal output and input
Control functions
Common
Monitor
signal output
Voltage monitor (VMON) Selectable monitor voltage range: 0 V to 5 V or 0 V to 10 V
Setting accuracy
2.5 % of f.s.
1
Current monitor (IMON) Selectable monitor voltage range: 0 V to 5 V or 0 V to 10 V
Setting accuracy
2.5 % of f.s.
1
Status
signal output
2
OUTON STATUS On when output is on.
CV STATUS Turns on during CV operation
CC STATUS Turns on during CC operation
ALARM STATUS Turns on when an alarm has been activated
POWER ON STATUS Turns on when the power is turned on
Trigger signal Input (TRG IN) Logic selectable: LOW (0 V to 1.5 V), HIGH (3.5 V to 5 V)
Input impedance: 10 kΩ (TYP)
Output (TRG OUT) Logic selectable: LOW (0 V to 0.6 V), HIGH (4.2 V to 5 V)
Pulse width: 100 µs (TYP)
1. f.s. is the full scale at the selected range. It is 10 V for the 10 V range and 5 V for the 5 V range.
2. Photocoupler open collector output;
maximum voltage 30 V, maximum current (sink) 8 mA; isolated from the output and control circuits; status
commons are floating (withstand voltage of less than or equal to 60 V); and status signals are not mutually
isolated.
Common
External
control
Output voltage control
(VPGM)
0 % to 100 % of the rated output voltage
Selectable control voltage range: 0 V to 5 V or 0 V to 10 V
Accuracy 5 % of rating
Output current control
(IPGM)
0 % to 100 % of the rated output current
Selectable control voltage range: 0 V to 5 V or 0 V to 10 V
Accuracy 5 % of rating
Output on/ off control
OUTPUT ON/OFF CONT
Logic selectable:
Output on when set to LOW (0 V to 0.5 V) or shorted; output off
when set to HIGH (4.5 V or 5 V) or open
Output on when set to HIGH (4.5 V to 5 V) or open; output off when
set to LOW (0 V or 0.5 V) or shorted
Output shutdown control
SHUT DOWN
Output on when set to LOW (0 V to 0.5 V) or shorted
Alarm clear control
ALM CLR
Alarm cleared when set to LOW (0 V to 0.5 V) or shorted