12.3.4 Programming Slave
The program updating proce
ss for the slave DSP can be started immediately after updating the
master DSP, no reset required.
• Select “Slave“ (4, Fig. 12-2) with the left-hand switch.
• Under File, select the *.ehx file for updating the slave DSP.
• Under Programming Configuration, make the settings shown.
• Under Port, select the serial interface (2).
• Click the command button Program (3).
• This starts the slave updating process.