1. After selecting Yes on the screen to the left below and pressing [OK/MUTE],
watch the Inlet Relief Valve and slowly reduce the remote exhaust system flow
until the Inlet Relief Valve begins to flutter, but is not completely closed. Hold the
remote exhaust system flow at this point. The state of the Inlet Relief Valve can
be seen on the display screen. The Inlet Relief Valve should still be Open. In
this state, press [OK/MUTE].
2. When the screen below is shown, the biosafety cabinet’s internal blowers will
shut off. Using a small, flat blade screwdriver, if the VSS status shown is Open,
adjust the VSS set screw slowly clockwise until the VSS status changes to
Closed. See Figure 9-4 for reference. Then slowly turn the set screw
counterclockwise until the status changes to Open. Because the screen is not
readily visible when adjusting the set screw, the biosafety cabinet will play an
audible alarm tone whenever the VSS status is Closed. Stop turning the set
screw counterclockwise when this tone stops. The VSS Calibration is complete.
3. Restore the remote exhaust system to nominal airflow. Press [OK/MUTE].