EasyManuals Logo

Labconco Logic+ A2 User Manual

Labconco Logic+ A2
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
91
QuickChart Type A2 5-ft & 6-ft
Type A2 Cabinet Width
5-ft
6-ft
Sash Opening (inches)
8
9
10
12
8
9
10
12
Starting Serial #
1905_
1905_
1905_
2006_
1905_
1905_
1905_
2006_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
60+/-5
55+/-5
55+/-5
55+/-5
60+/-5
Grid points (rows x columns)
30(3x10)
30(3x10)
30(3x10)
30(3x10)
36(3x12)
36(3x12)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
6.0
6.0
6.0
6.0
Distance between rows
5.75
5.75
5.55
5.44
5.75
5.75
5.55
5.44
Distance between columns
5.39
5.39
5.39
5.39
5.50
5.50
5.50
5.50
Inflow/Exhaust Data
Nominal Average Inflow (FPM)
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
105+/-5
Sash Open Area (Sq. Ft)
3.36
3.78
4.20
5.05
4.03
4.53
5.03
6.06
Nominal Avg. Inflow Vol. (CFM)
353
397
441
530
423
476
529
636
Avg. Inflow Vol. Range (CFM)
336-370
378-416
420-462
505-555
403-443
453-498
503-553
606-666
Secondary Inflow Data
Sash Opening Template
1
5+
5+
5+
5+
6+
6+
6+
6+
Sensor distance (inches)
2
3.25
3.25
3.25
3.25
4
4
4
4
# of Test points
10
10
10
10
12
12
12
12
Avg. Inflow Vel. Range (FPM)
336-370
378-416
420-462
505-556
220-242
248-272
275-302
303-333
Correction Factor (CF)
1
1
1
1
1.83
1.83
1.83
2.0
Corrected Inflow Volume (CIV)
(Avg. velocity x CF)
336-370
378-416
420-462
505-555
403-443
453-498
503-553
606-666
HEPA Filter Leak Test Data
Air Displacement (CFM)
900
950
1000
1135
1100
1140
1200
1380
Laskin Nozzles needed
2
2
2
3
2
2
2
3
Theoretical aerosol conc. (ug/l)
3
30
28
27
24
25
24
23
20
Actual aerosol conc. (ug/l)
4
17
14
14
13
17
15
15
10
Supply HEPA Data
Labconco P/N
3838402
3838402
3838402
3838402
3838403
3838403
3838403
3838403
Width x Depth x Height (in.)
5
60x18x3.06
60x18x3.06
60x18x3.06
60x18x3.06
72x18x3.06
72x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
555
555
555
555
665
665
665
665
Pressure Drop (in. H
2
O)
.5+/-.1”
.5+/-.1”
.5+/-.1”
.5+/-.1”
.38-.5”
.38-.5”
.38-.5”
.38-.5”
Exhaust HEPA Data
Labconco P/N
3838502
3838502
3838502
3838502
3838503
3838503
3838503
3838503
Width x Depth x Height (in.)
5
36x18x5.88
36x18x5.88
36x18x5.88
36x18x5.88
48x18x5.88
48x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
441
441
441
441
529
529
529
529
Performance (in. H
2
O)
.2-.37”
.2-.37”
.2-.37”
.2-.37”
.2-.37”
.2-.37”
.2-.37”
.2-.37”
LED Lamp Data (2 each)
7
Labconco P/N
1297505
1297505
1297505
1297505
1297506
1297506
1297506
1297506
LED Lamps (Direct Drive)
24T8-60
24T8-60
24T8-60
24T8-60
32T8-72
32T8-72
32T8-72
32T8-72
Color (
o
K)
4000
4000
4000
4000
4000
4000
4000
4000
Lumens
2400
2400
2400
2400
3200
3200
3200
3200
Glass Type
Frosted
Frosted
Frosted
Frosted
Frosted
Frosted
Frosted
Frosted
Ventus Canopy Data
Labconco Canopy P/N
3889202
3889202
3889202
3889202
3889203
3889203
3889203
3889203
Canopy Slot Area (ft
2
)
.12
.12
.12
.12
.12
.12
.12
.12
Nominal Canopy Inflow (CFM)
50
50
50
50
50
50
50
50
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
240-260
240-260
240-260
240-260
Canopy Vacuum (in. H
2
O)
10
0.15
0.15
0.15
0.15
0.15
0.15
0.15
0.15
DIM Total Exhaust Volume
Cabinet + Canopy (CFM)
8
403
447
491
580
473
526
579
686
Traverse Total Exhaust Volume
Cabinet + Canopy (CFM)
9
443
492
540
638
520
579
637
755
Volume of Hull
Cubic Feet
49
49
49
49
58
58
58
58
See footnotes on following page

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Logic+ A2 and is the answer not in the manual?

Labconco Logic+ A2 Specifications

General IconGeneral
BrandLabconco
ModelLogic+ A2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals