Change Configuration
Changes to the biosafety cabinet configuration can be made in the field. This typically
occurs after an accessory is field installed, or the user requests a change to the sash
height (between 8 and 10 inches) on a Type C1 only.
Never change the following configuration parameters:
Cabinet Type if the cabinet type is changed, it can result in incorrect operation
of the biosafety cabinet, and the biosafety cabinet may not protect the product,
personnel, or environment.
Sash Height On Type A2 and B2 cabinets, never change the sash height.
The following instructions detail all screens in the Configuration menu, which is
password protected. The password is [LIGHT] [UV LIGHT] [TIMER] [TIMER]
[OK/MUTE].
1. From the Main Menu, press [MENU], using [UP] or [DOWN] select the Tools
icon. Press [OK/MUTE].
2. Use [UP] or [DOWN] to select the Password icon, press [OK/MUTE].
[MENU]
Tools
Password
[OK/MUTE]
WARNING !
RESTRICTED AREA
FOR TECHNICIAN
USE ONLY