Change Configuration
Changes to the biosafety cabinet configuration can be made in the field. This typically
occurs after an accessory is field installed, or the user requests a change to the sash
height (between 8 and 10 inches) on a Type C1 only.
Never change the following configuration parameters:
Cabinet Type – if the cabinet type is changed, it can result in incorrect operation
of the biosafety cabinet, and the biosafety cabinet may not protect the product,
personnel, or environment.
Sash Height – On Type A2 and B2 cabinets, never change the sash height.
The following instructions detail all screens in the Configuration menu, which is
password protected. The password is [LIGHT] [UV LIGHT] [TIMER] [TIMER]
[OK/MUTE].
1. From the Main Menu, press [MENU], using [UP] or [DOWN] select the Tools
icon. Press [OK/MUTE].
2. Use [UP] or [DOWN] to select the Password icon, press [OK/MUTE].