187
21: Prefilter Kit
All Type A2 and B2 biosafety cabinets are prepared with the necessary mating
components to accept a Prefilter Kit. To install the Prefilter Kit, follow these instructions.
Kit Catalog Numbers (Type A2 and B2 only)
Installation
Decontaminate all interior work surfaces and work surface components before installing
the prefilter.
1. Remove the work surface of the cabinet by lifting the front edge by one of the
front corner knobs. Slide the work surface forward to disengage the rear tabs.
Support the underneath side with one hand while stabilizing the front edge with
the other. Set the work surface aside. Reference Figure 21-1.
2. Pinch and lift the work surface support straight up to remove it. Make sure it is
surface decontaminated before removing it from the biosafety cabinet.
Reference Figure 21-1.
3. Access the towel catch by pivoting the bottom out towards you. Surface
decontaminate the towel catch before removing it from the biosafety cabinet.
Reference Figure 21-1.
Cabinet
width (feet)
Catalog Number
G4 Paper
Aluminum
3-ft
3406000
3406100
4-ft
3406001
3406101
5-ft
3406002
3406102
6-ft
3406003
3406103
Table 21-1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Logic+ A2 and is the answer not in the manual?

Labconco Logic+ A2 Specifications

General IconGeneral
BrandLabconco
ModelLogic+ A2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals