169
Calibrate for Operation
After canopy installation, you MUST reconfigure the biosafety cabinet to have a canopy.
1. From the Home Screen, press the [MENU] button.
2. Press the [DOWN] button until the Tools option is
highlighted.
3. Press [OK/MUTE] to enter the menu screen.
4. Press the [DOWN] button until the Password option is
highlighted, then press [OK/MUTE] to select the
certifier’s submenus.
5. You will see a Warning Screen alerting you that you are
about to enter an area that can alter the BSC’s settings;
Press [OK/MUTE] to input the password.
6. When requested for the password press: [LIGHT],
[UV LIGHT], [TIMER], [TIMER] then [OK/MUTE].
7. If the password is entered correctly, you can select
between altering the cabinet’s configuration, or
calibration. Highlight the Configuration option, then
press [OK/MUTE] to start the configuration procedure.