170
8. This screen allows you to set the Type of cabinet.
Leave this setting alone, and press [OK/MUTE].
9. This screen allows you to configure the cabinet for an
airflow sensor or not. If you are not changing the
selection, press [OK/MUTE].
10. This screen configures the unit for a UV light or not.
Leave this setting alone, and press [OK/MUTE].
11. This screen allows you to configure the cabinet for a
canopy or not. Use [UP] or [DOWN] to select Canopy.
After your selection is chosen, press [OK/MUTE].
12. This screen allows you to configure the cabinet’s sash
height. Leave this setting alone, and press [OK/MUTE].
The unit will then return to the Home Screen in the new
configuration.
13. Take Inflow and Downflow readings as described in Section 9: Calibration. If not
within acceptable tolerance for the model, adjust the biosafety cabinet’s internal
damper and/or blower(s) speed until Inflow and Downflow are correct. Use the
original blower speed(s) as a reference.