35
2. The Info icon should be highlighted, if not select Info, press [OK/MUTE].
*Option(s) not installed are shown with a strikethrough. To activate options (after field
installation), see Change Configuration later in Section 8. To change the blower
speed(s), see Section 9: Calibration and Certification.
Type A2 12
-inch
and C1
Info
AFS
UV LIGHT
CANOPY
C1 - A MODE
REV. D
10 INCH
[OK/MUTE]
CURRENT / AVG = 1217 / 1219
NEW FILTER / MAX = 1074 / 1800
REV. D
SUPPLY
CURRENT / AVG = 1103 / 1104
NEW FILTER / MAX = 994 / 1800
REV. D
EXHAUST
[UP] [DOWN]
[UP] [DOWN]
Type
Sash
Height
Option(s)
Installed *
Software
Version
Blower
PWM
Only shown on two
blower models
(A2 12-inch and C1)
Instantaneous Blower RPM /
Time dampened average Blower RPM
Blower RPM with new HEPA filters /
Maximum Blower RPM
This screen only shown
on two blower models
(A2 12-inch and C1)