Optical Backscatter Reflectometer 4600 135
User Guide
match the new Integration Width (or Sensing Range) and
the
Integration
(or
Sensing
)
start set using the CONF:INTS
command.
CONFigure:INTegrationWidth?
Usage: CONF:INTW?
Description: Retrieves the OBR Integration Width in the units set by
CONF:XUNI (page 139).
[In optional Sensing mode (page 48), this returns the
Sensing Range, also in the units set by CONF:XUNI.
Alternatively, you may use the CONF:SRAN? command
(see below) to retrieve the Sensing Range.]
Response: Returns the OBR Integration Width (or Sensing Range)
in meters.
CONFigure:DECImation
Usage: CONF:DECI { number of decimated points }
Description: Sets the number of decimated points
Response: None.
Notes: This command controls the number of points that will be
returned by the FETC:XAXI? and FETC:MEAS?
commands, as well as the number of points written to the
text file using the SYST:SAVT command. It does not
affect the amount of data written to the binary file using
SYST:SAVE; the binary file always contains the full
measurement data.
Example: CONF:DECI 4096
This command sets the number of decimated points to 4096.
CONFigure:DECImation?
Usage: CONF:DECI?
Description: Retrieves the number of decimated points
Response: Returns the number of decimated points
CONFigure:DistanceRANge
Usage: CONF:DRAN {distanceRange}
Description: Sets the OBR Distance Range to Normal (0) or Extended