EasyManua.ls Logo

Luna 4600 - Page 152

Luna 4600
236 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
140 Chapter 7
Controlling the OBR Remotely
7
Description: Sets the gauge length in cm. This is the new name for
CONF:SRES {sensingResolution}. The CONF:SRES
command is still supported, but may be discontinued in the
future.
Response:
None
Examples: CONF:GLEN
1.5
Sets the gauge length to 1.5 cm.
CONFigure:GaugeLENngth?
Usage:
CONF:GLEN?
Description: Queries the gauge length. This is the new name for
CONF:SRES? {sensingResolution}. The CONF:SRES?
query will continue to be supported, but may be
discontinued in the future.
Response: Returns the gauge length in cm.
Example: CONF:GLEN?
This query returns the gauge length in cm.
CONFigure:SensorSPAcing
Usage:
CONF:SSPA
Description: Sets the Sensor Spacing in cm. The sensor spacing was
previously tied to the Sensing Resolution value, and both
were set by the
CONF:SRES
command. It is now a separate
value with its own command.
Response:
None.
Example: CONF:SSPA
2.0
Sets the Sensor Spacing to 2.0 cm.
CONFigure:SensorSPAcing?
Usage:
CONF:SSPA?
Description: Queries the Sensor Spacing.
Response: Returns the Sensor Spacing in cm.
Example: CONF:SSPA?
This query returns the sensor spacing in cm.

Table of Contents