Optical Backscatter Reflectometer 4600 141
User Guide
CONFigure:SensingRESolution and CONF:SRES?
Note: Being replaced by the command CONF:GLEN and
CONF:SSPA (above).
CONFigure:XaxisUNIts
Usage: CONF:XUNI {units}
Description: Sets the X-axis units being used to specify the Integration
(or Sensing) Area. These are the units that will be used by
the CONF:INTW, CONF:INTS, and CONF:INTC
commands and queries to report or set the Integration
Width (or Sensing Range), start, and center. The units
parameter must have one of the following values:
0 : nanoseconds
1 : meters
2 : feet
3 : inches
4 : millimeters
Response: None
Example: CONF:XUNI 0
CONF:INTW?
Response: 15.010384
The CONF:XUNI command sets the x-axis units to
nanoseconds. This means that the CONF:INTW? response
should be interpreted as 15.01 nanoseconds, and that any
CONF:INTW, CONF:INTS, or CONF:INTC commands
should specify their units in nanoseconds.
CONFigure:XaxisUNIts?
Usage: CONF:XUNI?
Description: Queries the X-axis units being used to specify the
Integration
Width (or Sensing Range). These are the units
that will be used by the CONF:INTW, CONF:INTS, and
CONF:INTC commands and queries to report or set the
Integration (or Sensing) width, start and center.