16 Chapter 2
Assembly and Startup
2
17 Start the control software by double-clicking the OBR desktop
icon or by selecting OBR Software from the programs group in the
Windows
®
Start menu. An initial window will appear for a few moments
while the user interface is initialized.
If the control software is started when the instrument is not
powered on, a communication failure error message
opens. However, the software can be used in desktop
I
m
port
a
n
t
analysis mode when the instrument is not powered by
clicking OK in the message box.
To operate in normal mode (which is required for
scanning), turn on the instrument and then restart the OBR
software.
Once the control software is running, the OBR is ready to Acquire Reference. (See
“Instrument Configuration” on page 34, and “Calibrating the System” on page 35.)
Any time the equipment stabilizes to a new ambient temperature (anywhere between
10–40 ºC) the user may wish to perform another Reference Scan for greater
accuracy.