Figure 87 CT Burden Test Connection Diagram
14. Check all connections to ensure correctness
15. Hit Blue Play button to begin test.
.
16. The Burden test will begin.
Voltage (V): The MVCT displays the real time voltage readings of the burden while the test is in
progress.
Current (A): The MVCT the real time current readings for the burden while the test is in
progress.
17. At any point of time Test can be terminated by hitting the ABORT button at the top
of the screen.