65
Settings that are accessible will be in black. VT accessible buttons shown below in black and a
description of what the user can change follows:
Standard: Select between ANSI, and IEC
Color Options: Set colors in excitation curves
Tolerance Settings: Not Available
Ratio Error on Primary/Secondary Current: Not Available
Default Ratio/Phase Error Table: Not Available
Report and Assessment: Allows the user to customize the report. He can select whether
or not to display the ratio and phase error tables
Set Date Time Change the time and date stored in the MVCT. This is only available in
MVCT with integrated display or if they are controlled with an STVI
Number Format Select between U.S. and International number formatting. This is only
available in MVCT with integrated display or if they are controlled with an STVI
Show Connection Diagrams: Allows user to turn off the functionality showing connection
diagrams before each test. This is not recommended because of safety concerns.
Knee Prediction Enabled/Disabled: Not Available
Lead Check Enable/Disable Allows the user to enable and disable the lead checking
algorithms. If enabled the MVCT will check connections prior to a test and monitor voltages and
currents during the test. Anytime an unexpected value is detected the user will be warned to
check the connections.
Nameplate Estimate Enable/Disable: Not Available
Default Number of Taps: Not Available
Use Binary Input as Stop: Not Available
Default Settings Options: Save changes made to defaults, restore defaults, or restore
factory defaults
Upgrade Firmware: Upgrade firmware for the MVCT or the STVI
Display Versions: View all Software and hardware versions of the device.
Languages: Select between American English, International English, Spanish, French, and
German, Turkish
Primary Noise Filtration Enable/Disable Allows the user to enable and disable the primary
noise filtration algorithm. This should only be enabled in environments where induced voltage
is effecting the testing results
Logging: Enable logging, clear all log files, or copy log files to USB. . This is only available
in MVCT with integrated display or if they are controlled with an STVI