Page 55
Serial Interface RS422
optoNCDT 1402
Error Output (Analog output) 8.3.11
Name: SET_ANALOG_ERROR_HANDLER
Description: Hold or not hold last measurement value.
Parameter:
X = 0 --> hold last measurement value -
X = 1 --> error value (3.75 mA) -
X = 2 … 99 --> hold last measurement value for 2 … 99 images respectively cycles -
This command only affects the analog output. If set to X = 0, the last valid measurement value will continue
to be issued if an error occurs (no object, invalid object, object outside the measurement range or laser
turned off). If set to X = 1, an error signal will be generated for the current output that has an error value of
3.75 mA. If set to X = 2 ... 99, the last valid measurement value will continue to be issued for X measuring
cycles before an error signal is generated on the analog output.
Format: 31 24 23 16 15 8 7 0 hex
„+“ „+“ „+“ 0x0d („CR“) 0x2B2B2B0D
„l“ „L“ „D“ „1“ 0x494C4431
0x20 0x81 0x00 0x03 0x20810003
0x00 0x00 0x00 0x0X 0x0000000X
Reply: 31 24 23 16 15 8 7 0 hex
„l“ „L“ „D“ „1“ 0x494C4431
0xA0 0x81 0x00 0x02 0xA0810002
0x20 0x20 0x0D 0x0A 0x20200D0A
Default setting:
error value