EasyManuals Logo

MICRO-EPSILON ILD1402-5 User Manual

MICRO-EPSILON ILD1402-5
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Page 45
Serial Interface RS422
optoNCDT 1402
Memory mode
0x20F70003 SET_SAVE_SETTINGS_MODE Volatile / nonvolatile
Scaling values
0x20F90004 SET_TEACH_VALUE Sets T1 + T2 0 ... 16368
0x20FA0002 RESET_TEACH_VALUE Sets T1 = 0 / T2 = 16368
Search algorithm
0x20FB0003 SET_PEAKSEARCHING First peak, last peak, global maximum
Threshold
0x20FC0003 SET_THRESHOLD
Lower than standard, standard, higher than standard,
highest
Switch data protocol
0x20F20002 SET_CIMODE_1401 Sensor operates with ILD1401 data protocol
0x2D2D2D52 GET_CI_MODE Requests the command interpreter state of the sensor
Reading the Sensor Parameters8.3.3
Name: Get_Info
Description: Supplies the info string. This shows all parameters currently stored in the sensor.
Format: 31 24 23 16 15 8 7 0 hex
„+“ „+“ „+“ 0x0d („CR“) 0x2B2B2B0D
„l“ „L“ „D“ „1“ 0x494C4431
0x20 0x49 0x00 0x02 0x20490002