Page 50
Serial Interface RS422
optoNCDT 1402
Peak searching
0x00 0x00 0x00 0x0X 0x0000000X
Threshold
0x00 0x00 0x00 0x0X 0x0000000X
Measuring range
0x00 0x00 0xXX 0xXX 0x0000XXXX
Reserved 1
0x00 0x00 0x00 0x0X 0x00000000
Reserved 2
0x00 0x00 0xXX 0xXX 0x00000000
Reserved 3
0x00 0x00 0x00 0x0X 0x00000000
Reserved 4
0x00 0x00 0x00 0x0X 0x00000000
Last data word
0x20 0x20 0x0D 0x0A 0x20200D0A
Average Type and Average Number8.3.5
Name: SET_AV
Description: Sets the average type and the average number N.
Parameter:
Average type -
X = 0 --> Moving average
X = 1 --> Median
Average number -
XX = 1 ... 128 --> Moving average, if average type = moving average
XX = 3, 5, 7, 9 --> Median, if average type = Median
Default setting:
moving average 1, thus
no averaging