Product Specifications
A-14
A.5.4 NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1,
EN1060-3 and SP10
Technique Oscillometric
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 or 480 min
STAT mode cycle time 5 min
Adult, pediatric: 180 s
Max measurement time
Neonate: 90 s
Heart rate range 40 to 240 bpm
Adult Pediatric Neonate
Systolic: 40 to 270 40 to 200 40 to 135
Diastolic: 10 to 210 10 to 150 10 to 100
Measurement ranges
(mmHg)
Mean: 20 to 230 20 to 165 20 to 110
Accuracy
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure
Adult:
Pediatric:
Neonate:
178±5 mmHg
133±10 mmHg
87±5 mmHg
Software overpressure
protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
Hardware overpressure
protection
Adult:
Pediatric:
Neonate:
330±5 mmHg
330±5 mmHg
165±5 mmHg